Q-Plex™ Human Pneumococcal IgG (23-Plex)

Q-Plex Human Pneumococcal IgG (23-Plex)
List of Serotypes: 1, 2, 3, 4, 5, 6B, 7F, 8, 9N, 9V, 10A, 11A, 12F, 14, 15B, 17F, 18C, 19A, 19F, 20, 22F, 23F, 33F
- Species Specificity: Human
- Validated Sample Type: Serum
- Sample Volume Required: 0.25 µL
- Estimated Assay Length: Less than 1.25 hours
- Intra-Assay Cross Reactivity: No significant interference observed. <1% cross-reactivity observed.
- Assay Type: Indirect
- Detection Method: Chemiluminescent
- Multiplex Format: 96-well solid plate
- Catalog 23-Plex: #481149HU
- Catalog 14-Plex: #481749HU
- Catalog 9-Plex: #481649HU
Q-Plex Human Pneumococcal IgG (23-Plex)
Serotypes: 1, 2, 3, 4, 5, 6B, 7F, 8, 9N, 9V, 10A, 11A, 12F, 14, 15B, 17F, 18C, 19A, 19F, 20, 22F, 23F, 33F
This kit is a chemiluminescent multiplex ELISA run on two 96-well plates, one plate will run 14 serotypes (1, 3, 4, 5, 6B, 7F, 8, 9N, 9V, 12F, 14, 18C, 19F, 23F) and the other plate will run the remaining 9 serotypes (2, 10A, 11A, 15B, 17F, 19A, 20, 22F, 33F).
Q-Plex Human Pneumococcal IgG (14-Plex)
Serotypes: 1, 3, 4, 5, 6B, 7F, 8, 9N, 9V, 12F, 14, 18C, 19F, 23F
This kit can be ordered individually from the 23-Plex.
Q-Plex Human Pneumococcal IgG (9-Plex)
Serotypes: 2, 10A, 11A, 15B, 17F, 19A, 20, 22F, 33F
This kit can be ordered individually from the 23-Plex.
Q-Plex™ Human Pneumococcal IgG (14-Plex)
Serotype | Assay Type | Calibrator Range | Upper Limit ofQuantification (ULOQ) | Lower Limit ofQuantification (LLOQ) | Limit of Detection | Precision* (Inter-assay) | Precision* (Intra-assay) | Average Linearity |
---|---|---|---|---|---|---|---|---|
1 | Indirect | 0.85 - 0.00117 (μg/mL) | 0.259 (μg/mL) | 0.000468 (μg/mL) | 0.000583 (μg/mL) | 6% | 7% | 97% |
3 | Indirect | 0.0145 - 0.00020 (μg/mL) | 0.294 (μg/mL) | 0.000533 (μg/mL) | 9.95E-05 (μg/mL) | 12% | 14% | 99% |
4 | Indirect | 0.33 - 0.00045 (μg/mL) | 0.154 (μg/mL) | 0.000273 (μg/mL) | 0.000226 (μg/mL) | 5% | 6% | 99% |
5 | Indirect | 0.75 - 0.00103 (μg/mL) | 0.371 (μg/mL) | 0.000676 (μg/mL) | 0.000514 (μg/mL) | 8% | 8% | 102% |
6B | Indirect | 0.91 - 0.00125 (μg/mL) | 0.133 (μg/mL) | 0.000247 (μg/mL) | 0.000624 (μg/mL) | 8% | 9% | 103% |
7F | Indirect | 0.83 - 0.00114 (μg/mL) | 0.392 (μg/mL) | 0.00286 (μg/mL) | 0.000569 (μg/mL) | 6% | 7% | 100% |
8 | Indirect | 1.42 - 0.00195 (μg/mL) | 0.133 (μg/mL) | 0.000962 (μg/mL) | 0.000974 (μg/mL) | 7% | 8% | 97% |
9N | Indirect | 0.87 - 0.00119 (μg/mL) | 0.154 (μg/mL) | 0.001105 (μg/mL) | 0.000597 (μg/mL) | 8% | 10% | 102% |
9V | Indirect | 0.64 - 0.00088 (μg/mL) | 0.203 (μg/mL) | 0.00143 (μg/mL) | 0.000439 (μg/mL) | 8% | 8% | 102% |
12F | Indirect | 0.22 - 0.00030 (μg/mL) | 0.07 (μg/mL) | 0.00052 (μg/mL) | 0.000151 (μg/mL) | 11% | 9% | 93% |
14 | Indirect | 3.8 - 0.00521 (μg/mL) | 0.119 (μg/mL) | 0.000871 (μg/mL) | 0.002606 (μg/mL) | 7% | 7% | 98% |
18C | Indirect | 0.73 - 0.00100 (μg/mL) | 0.217 (μg/mL) | 0.00156 (μg/mL) | 0.000501 (μg/mL) | 5% | 5% | 100% |
19F | Indirect | 1.46 - 0.00200 (μg/mL) | 0.924 (μg/mL) | 0.00676 (μg/mL) | 0.001001 (μg/mL) | 9% | 12% | 100% |
23F | Indirect | 0.6 - 0.00082 (μg/mL) | 0.084 (μg/mL) | 0.000624 (μg/mL) | 0.000412 (μg/mL) | 10% | 10% | 99% |
Q-Plex™ Human Pneumococcal IgG (9-Plex)
Serotype | Assay Type | Calibrator Range | Upper Limit ofQuantification (ULOQ) | Lower Limit ofQuantification (LLOQ) | Limit of Detection | Precision* (Inter-assay) | Precision* (Intra-assay) | Average Linearity |
---|---|---|---|---|---|---|---|---|
2 | Indirect | 2.46 - 0.00337 (μg/mL) | 2.44 (μg/mL) | 0.00061 (μg/mL) | 0.00169 (μg/mL) | 10% | 12% | 101% |
10A | Indirect | 1.29 - 0.00177 (μg/mL) | 0.3 (μg/mL) | 0.00032 (μg/mL) | 0.000885 (μg/mL) | 7% | 8% | 106% |
11A | Indirect | 0.508 - 0.00070 (μg/mL) | 0.14 (μg/mL) | 0.00013 (μg/mL) | 0.000348 (μg/mL) | 7% | 8% | 103% |
15B | Indirect | 1.69 - 0.00232 (μg/mL) | 0.47 (μg/mL) | 0.00044 (μg/mL) | 0.001159 | 9% | 10% | 105% |
17F | Indirect | 0.85 - 0.00117 (μg/mL) | 0.79 (μg/mL) | 0.00067 (μg/mL) | 0.000583 (μg/mL) | 9% | 8% | 98% |
19A | Indirect | 1.39 - 0.00191 (μg/mL) | 1.24 (μg/mL) | 0.0061 (μg/mL) | 0.000953 (μg/mL) | 6% | 6% | 101% |
20 | Indirect | 1.05 - 0.00144 (μg/mL) | 0.26 (μg/mL) | 0.00028 (μg/mL) | 0.00072 (μg/mL) | 6% | 6% | 102% |
22F | Indirect | 0.95 - 0.00130 (μg/mL) | 0.95 (μg/mL) | 0.00094 (μg/mL) | 0.000652 (μg/mL) | 9% | 12% | 107% |
33F | Indirect | 1.07 - 0.00147 (μg/mL) | 0.27 (μg/mL) | 0.00026 (μg/mL) | 0.000734 (μg/mL) | 6% | 6% | 101% |
Background
Principle of the Assay

The Q-Plex™ Human Pneumococcal IgG (23-Plex) is a chemiluminescent multiplex ELISA run on two 96-well plates comprising the serotypes found in the Pneumovax®23 vaccine.
The Q-Plex™ Human Pneumococcal IgG (14-Plex) and Q-Plex™ Human Pneumococcal IgG (9-Plex) are also available for individual purchase.
Kit Content
Reagents Include:
- Detection Mix
- Calibrators
- Substrates
- Sample Diluents
- Streptavidin HRP
- Wash Buffer
Processing
Q-Plex™ Arrays are developed and optimized to work with Q-View™ Imagers. Quansys does not guarantee the results obtained from other imagers and not all imagers are compatible with Quansys Q-Plex Arrays.
Order Your Kit Today!
Not Finding What You Need?
Customize a Kit
Not seeing the biomarker or analyte you need? Customize your kit! The Q-Plex™ Custom Assays are fully quantitative ELISA-based Chemiluminescent assay kits allowing the concurrent measurement of up to 18 biomarkers or analytes.
Send in Your Samples
Choose up to 18 different biomarkers that will work best for your research. Send in your samples and let us run them for you.
Contact a Representative
Have questions about the kit or interested in more information? Contact one of our representative to get all your questions answered.
As researchers and problem solvers ourselves, we understand the value of sound data. We are proud to be a part of research that can better the world. We value the relationships, partnerships, and friendships that we have built with the people who trust and use our technology. We are committed to building these relationships. You can count on us to answer the phone and take time to thoroughly address questions or concerns about any of our products. In an industry that is reputed for grandiose claims, we trust that our quality standards and our customer service set us apart from the competition. If you’re happy with our products or if you think we can do something better, we hope you will let us know.